Quantitative Fluorescence Spectral Corrections, Inner Filter Effect Quantum Yield | PHYS 552, Lab Reports of Optics

Material Type: Lab; Class: Optical Spectroscopy; Subject: Physics; University: University of Illinois - Urbana-Champaign; Term: Unknown 1989;

Typology: Lab Reports

Pre 2010

Uploaded on 03/16/2009

koofers-user-pzt
koofers-user-pzt 🇺🇸

9 documents

1 / 14

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
Physics 552 Optical Spectroscopy (Fall 08)
- 1 -
Lab 4: Quantitative Fluorescence
Spectral Corrections, the Inner Filter Effect, and Quantum
Yield
In this week’s lab, we will use a fluorometer in Prof. Clegg’s lab to study some issues involved
with fluorescence measurements. The schematic design of a fluorometer is included below. This
particular one has two emission monochromators (right and left sides – we will be using the right
side monochromator).
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe

Partial preview of the text

Download Quantitative Fluorescence Spectral Corrections, Inner Filter Effect Quantum Yield | PHYS 552 and more Lab Reports Optics in PDF only on Docsity!

Lab 4: Quantitative Fluorescence

Spectral Corrections, the Inner Filter Effect, and Quantum

Yield

In this week’s lab, we will use a fluorometer in Prof. Clegg’s lab to study some issues involved with fluorescence measurements. The schematic design of a fluorometer is included below. This particular one has two emission monochromators (right and left sides – we will be using the right side monochromator).

Experiment I: Spectral corrections and the inner filter effect

(Fluorescein and Phenolphthalein)

Introduction – Spectral Corrections

When making quantitative fluorescence measurements, it is important to correct the measured spectra to account for losses throughout the system. Spectral corrections come in three forms: (1) Instrumental corrections (a) Excitation-side corrections (b) Emission-side corrections (2) Sample corrections

Instrumental Corrections Instrumental corrections on the excitation side are handled by the reference detector. This is done by splitting off a known fraction of the beam incident on the sample and measuring its intensity I 0 (λ) as a function of wavelength. The detected fluorescence Id(λ) is then normalized by dividing the measured intensity by the incident intensity at each wavelength. In effect, this assumes that the sample is excited with uniform, white-light excitation. It accounts for any variations or fluctuations in lamp power, as well as losses in the excitation monochromator. The excitation-side corrections are generally handled by the instrument by selecting to record in ratio or signal/reference mode.

The emission-side corrections , while similar in origin, must be handled differently by the instrument. In this case, we wish to know the probability p(λ) that an emitted photon will be successfully detected. Knowing this, we can back- track from the measured intensity to determine the fluorescence intensity I(λ) emitted by the sample. This correction will account for all losses in traveling from sample to detector including chromatic aberrations in the optics and the monochromator and detector efficiencies. These correction values are just the inverse of the relative probability of reaching the detector and triggering a response. They depend only on the emission wavelength λem. If the instrument has polarizers, it would also depend on the polarization of the emitted light. The correction is applied by multiplying the measured intensity at each wavelength by the appropriate value. In practice, the emission-side correction(s) are determining using either a calibrated lamp source or a series of known fluorophores with well-established emission spectra. In either case, the measured spectra (also known as technical spectra) are directly compared to the known source data to determine the probability curve.

We will measure the raw excitation and emission photon counts for the fluorescence spectra. Then, in your analysis, you will make the excitation side instrumental correction by dividing out the lamp excitation intensity measured by the reference detector. Then you will account for the

Fig. 1: Emission side spectral correction

  1. Turn off the photomultiplier tube (center position) and open up the right emission side compartment to check that you have 1 mm slits inserted. Close the compartment, then turn on the photomultiplier tube (push down).
  2. Start first with the pure fluorescein sample (G). Place the cuvette in the sample space on the sample turret (remember the orientation of the cuvette so that you can put it in the same way for all samples). Then click on the green play button to record the fluorescence spectra. When it is done, click yes to save the spectra to file. Repeat for the other 7 samples, A-F and the 0.1 N NaOH (buffer) sample. Record the peak wavelength and intensity in the table on the next page.
  1. Excitation Spectrum So you can get an idea of the necessity of excitation side corrections, we will take an excitation spectrum of fluorescein. a.) Place the G (pure fluorescein) sample once again in the fluorometer b.) Select the excitation spectrum experiment: >> Experiment>>User Defined>>#0Lambert_FluoresceinPhenolExc You will set the emission monochromator at 515 nm and scan the excitation monochromator from 400 to 510 nm. c.) Set the view to emission (or all): >> View>>Visualization In the dropbox next to y, choose either emission or all. d.) Save the excitation spectrum.
  2. You will also need to measure the absorption spectra. We will need these to correct for the inner-filter effect to recover the proper emission spectrum for fluorescein. The equations to use for correcting the emission spectra are derived in the appendix at the end of this write-up.

Use the Agilent 8453 absorption spectrometer you’re familiar with to measure the absorption spectra. At a minimum, the spectra should include the excitation wavelength and the full range of wavelengths you used for the emission spectra. Make sure you place the cuvettes in the cuvette holder in the same orientation for each sample. (Also be aware that the cuvette holder tends to leave a mark along one side of the cuvette so it is best to keep track of this.) Save a spectrum on the computer. Remember to blank the sample to the solvent 0.1N NaOH. Record your data in the following table:

absorption fluorescence sample λmax [nm] A (^) max λem [nm] I (^) em [A.U.] 0.1N NaOH -- G F E D C B A

sample % glycerol λabs [nm] A (^) max A 620 λem [nm] I (^) em [A.U.] Cy5A 0 Cy5B 20 Cy5C 40 Cy5D 60

  1. Using this data, we can now calculate the fluorescence quantum yield for each sample. By definition, the quantum yield is the ratio of the amount of output (fluorescence) to the amount of input (absorption):

The numerator is proportional to the total fluorescence signal coming out of the sample and the denominator is proportional to the total absorption of the sample. You will calculate the relative quantum yield from:

where I (^) F,em,peak is the fluorescence intensity at the peak of the emission spectrum and Aλ,ex is the absorbance at the excitation wavelength.

To get the absolute value of the quantum yield, we will use the known value of 0.27 for the quantum yield of Cy5 in aqueous solution (0% glycerol). The formula to calculate the absolute yield based on this is:

where n is the index of refraction of the solvent. You will need the following data:

% glycerol n viscosity (cP) 0 % 1.3330 1. 20 % 1.3663 2. 40 % 1.39768 6. 60 % 1.426 21.

photons absorbed

photonsemitted F

ex

rel Fempeak F A

I

,

, , λ

rel

rel

n

n

2

1 2 2

2 1 2

1 φ

φ φ

φ = ⋅

Appendix: Correction of fluorescence spectra for the inner-filter effect.

This note summarizes the derivation of the corrections for the inner-filter effect. The figure at the end shows the light paths under consideration as they travel through the cuvette / sample. The points labeled 1 to 5 (in the ovals) correspond to steps 1 to 5 below. The remaining steps are manipulations of these equations.

  1. Light of wavelength λx and intensity I 0 (photons/area) is incident on the front face of the cuvette. It is confined by the excitation optics to the dimensions Δy by Δz. The total number of photons is N 0 = I 0 Δy Δz. Over the infinitesimal area dA = dy dz , the number of incident photons is:

d^2 N 0 = I 0 dy dz [1]

  1. The light travels the distance x through the sample and may be absorbed by any compounds present. The total absorption (base- e ) per unit length is α(λ). At the excitation wavelength, the total absorption is α(λx ) and the shorthand notation is αx. The number of photons Nx transmitted the distance x is (Beer-Lambert law):

d^2 Nx = d^2 N 0 exp ( − α xx ) [2]

  1. Now consider the infinitesimal volume dx dy dz at the point (x,y,z) within the cuvette. We will calculate the amount of light absorbed by the fluorophore within this volume. The fluorophore has an absorption of αF at the excitation wavelength. Over the length dx , the number of photons absorbed is (differential form of Beer-Lambert law):

d^3 NA = d^2 Nx α F dx [3]

  1. The excited molecules have some probability qF (the quantum yield of fluorescence) of emitting the absorbed light at a longer wavelength λm ( m stands for e m ission). The remainder of the energy is lost as heat to the solvent. The fluorescence is emitted in all directions, but only a small amount of this is headed in the proper direction to reach the detector. In the figure, only the light traveling straight downward in the negative - y direction can reach the detector. In general, each point (x,y,z) within the sample will have some probability p(x,y,z) that the emitted photons will be detected. The number of fluorescent photons NF that have a chance to reach the detector is then given by:

d^3 NF = d^3 NAqF p ( x , y , z ) [4]

The probability p(x,y,z) is a function of the emission optics. In the first approximation, it is equal to the fraction of the solid angle striking the first optic (lens or mirror) collecting the emission. This probability does not include the corrections for the monochromator and detector efficiency (handled by the software) or the sample corrections (which we are treating explicitly in this calculation). In principle, we might determine p(x,y,z) using ASAP to model the fluorimeter hardware.

Note the identical form of the corrections for the excitation and emission wavelengths. All that is required to apply these corrections is information about the cuvette dimensions ( Li ), the detected volume ( Δi ), and the corrected absorption spectrum for the sample ( αi ).

  1. Equation [9] can be re-written in the form Nd^0 = Nd βx βm with the correction factors βx and βm corresponding to the e x citation and e m ission wavelengths / paths, respectively:

sinh

exp

i i

i i i i i

L

α

α α β (^) [10]

  1. If the fluorescence is collected from the full length of the cuvette ( Δi = Li ), then the correction in equation [10] has the form:

( i i )

i i i

L

L

1 exp

[11]

  1. Now consider a single-component system ( αx = αF ), a small detection volume ( Δx << 1 ), and negligible absorption at the emission wavelengths ( αm << 1 ). In this case, the measured signal given by equation [7] reduces to:

d #^0 Ins F x y z F exp F x

L

N I q

β α [12]

At low concentrations ( αF << 1 ), Nd#^ is directly proportional to the concentration of fluorophore (this must be true). At high concentrations ( αF >> 1 ), Nd#^ goes to zero (all the light is absorbed before reaching the center of the cuvette). Thus, somewhere between these end points, the detected signal must have a maximum value.

  1. To determine the absorption that gives the maximum detected signal, we take the derivative of equation [12] with respect to the fluorophore absorption αF , set it equal to zero, and solve to give:

x

F

L

max 2

α = [13]

  1. For a 1 cm cuvette, the maximum base-10 absorption coefficient is:

ln( 10 )

max max

AF = F ≈

[14]

LAMP

DETECTOR

1_d^2 N (^0) 3_d 3 N 2_d^2 N (^) x A

4_d 3 N (^) F

5_d^3 N (^) d

absorbed transmitted d 2 Nx+dx

0 x x+dx Lx

0

Ly

y

y+dy

ΔY

ΔX

N 0 = I 0 Δ y Δ z

Nd = Id Δ x Δ z

Detected fluorescence

confined to the volume

Δ x by Δ y by Δ z about the

center of the cuvette

LAMPLAMP

DETECTORDETECTOR

1_d^2 N (^0) 3_d 3 N 2_d^2 N (^) x 3_d 3 NA (^) A

4_d 3 N (^) F

5_d^3 N (^) d

absorbed transmitted dd 2 2 NNx+dxx+dx

0 xx x+dxx+dx Lx

0

Ly

y

y+dyy+dy

ΔY

ΔX

NN 00 = I= I 00 ΔΔ yy ΔΔ zz

NNdd = I= Idd ΔΔ xx ΔΔ zz

Detected fluorescence

confined to the volume

Δ x by Δ y by Δ z about the

center of the cuvette

Emission Wvl Excitation Emission ExcitationStdError EmissionStdErr 480 54024.9 2675.2 148.1 87. 482 53927.8 1747.3 233.63 45. 484 53823.9 2290.3 271.02 37. 486 54448.4 3267.6 265.13 40. . . where excitation is the number of counts at the reference photomultiplier tube, emission is the number of counts at the emission photomultiplier tube. Whether you’re measuring an excitation or an emission wavelength, your raw spectrum is given by the emission column.

For each spectra, you need to do the following:

  1. Do the excitation side correction by dividing the emission column by the excitation column.
  2. Now do the emission side correction by multiplying this new column by correction file provided. Note that the correction file is for a scan range larger than the range over which you took data, so you should adjust your correction file accordingly.
  3. Sample corrections a. Background fluorescence of the buffer : When taking fluorescence spectra, you should take spectra of blank samples (sample minus the fluorophore). This allows you to account for background fluorescence, the presence of fluorescent impurities in your solvent, as well as the presence of Rayleigh and Raman scattering. You should have done corrections 1 and 2 for the emission spectrum you took of the 0.1N NaOH buffer. Now for all other emission spectra for samples A to G, subtract out the spectrum of the NaOH buffer. (For our case, this has minimal effect). b. Inner filter effect. Now correct for the inner filter effects using Eq. 9 from the lab handout. Since we derived the equations for absorbance in base e but we measured the absorbance in base 10, we need to rewrite Eq.9 in base 10:

( ) ( ) 2 2

ln(10) ( ) ln(10)^ (^ ) 2 2 10 10 ln(10) ( ) ln(10) ( ) sinh (^) sinh (^2 )

ex em

ex x em^ y A A x y in fil corr ex x em y x (^) y

A A L L Em Em A A L (^) L

λ λ

λ λ

λ λ

⎛ ⎞ ⎛ ⎞ ⎜⎝ ⎟⎠ ⎜⎝ ⎟⎠ − −

⎛ (^) Δ ⎞⎛^ Δ ⎞ ⎜ ⎟ ⎜^ ⎟ ⎜ ⎟⎜^ ⎟ = (^) ⎜ ⎟ ⎜ (^) ⎛ (^) Δ ⎞ ⎟ ⎛ (^) Δ ⎞ ⎜ (^) ⎜ ⎟ ⎟⎜^ ⎜ ⎟⎟ ⎜ ⎟ (^) ⎜ ⎜ ⎟⎟ ⎝ ⎝^ ⎠⎠ (^) ⎝ ⎝ ⎠⎠ where Δx and Δy are from the empirically determined detection volumes for your fluorometer. For our case, Δx = 1 cm and Δy = 0.1 cm works well. A(λex ) is the absorption at the excitation wavelength 470nm and A(λem) is the absorption at the emission wavelength you are scanning (i.e, for an emission spectrum, you need A(λem) is a function of the emission wavelength you are scanning. Note that your emission spectra start at 480 nm but the absorption spectrum starts at a lower wavelength. You should truncate off the data for the appropriate data points when you are doing your correction (although you’ll need to use the OD at 470 nm for A(λex ).

  1. Turn in a plot of the uncorrected emission spectra. Use one plot for all the samples.
  1. Turn in a plot of the emission spectra with the instrumental corrections. Use one plot for all the samples.

  2. Turn in a plot of the emission spectra with instrumental and sample corrections. Use one plot for all the samples.

  3. The fluorescein concentrations are the same in all 6 samples. Make a plot of the raw and the corrected emission peak value vs. phenolphthalein concentration (just use the OD at the absorption peak of phenolphthalein around 550 nm as the x-axis). How good are your corrections?

7) Excitation spectrum of fluorescein (G). Here, we’ll just use this excitation spectrum to take a look at the effect of excitation side corrections. Make a plot of the raw excitation spectrum, and the excitation spectrum where you make excitation side correction by dividing Em by Ex. Put both spectra on the same plot (either make a left and right axis plot or renormalize one of the spectra).

  1. Show your work. You can do this one of two ways. a) Email your analysis file to [email protected] with the subject heading P552 Lab4 Analysis. If you used Excel, you can just email the Excel file. For all other programs please make an ascii file of the worksheet or equivalent containing your raw and corrected spectra. Make sure we can understand the manipulations you did to correct the spectrum. I.e., all parameters used and column manipulations should be stated. Or, b)Print out the first page of your worksheet showing all columns (please don’t print all rows). Then write down the manipulations you did and parameters you used on the printout.

Experiment II: Quantum yield calculations (Cy5 and Glycerol)

Follow the instructions in the experiment handout to calculate the quantum yield. For our purpose here, the only correction you need to do for the fluorescence spectra is to divide the Emission column by the excitation column to get Iem.

9). Calculate the absolute quantum yields for the Cy5 samples and prepare a plot versus the solvent viscosity. Present your table of data from p.7 of the experiment handout. Briefly explain why the quantum yield increases with increasing viscosity.

10). The factor of n^2 in the quantum yield calculation arises because the fluorescent rays are refracted as they exit the sample into air on their way toward the detector. Briefly explain why a larger index solvent leads to less light reaching the detector (all other things being equal). This is easy to derive using Snell’s Law, but I am not asking for the full derivation.

11). In the calculation of quantum yield, is it correct to use only the peak emission intensity? What about all the other photons of different wavelengths? What should actually be used in this calculation? Why were we able to use just the peak intensity in this case?