Donald Peck, PhD
Henry Ford Health System
Detroit, Michigan USA
Radiation Dose Structured Report
Current and Future
1
pf3
pf4
pf5
pf8
pf9
pfa
pfd
pfe
pff
pf12
pf13
pf14
pf15
pf16
pf17
pf18
pf19
pf1a

Partial preview of the text

Download Radiation Dose Structured Report and more Exams Radiography in PDF only on Docsity!

Donald Peck, PhD

Henry Ford Health System

Detroit, Michigan USA

Radiation Dose Structured Report

Current and Future

Radiation Dose SR

l Original developed in 2005 for

Projection Radiography

  • Added 3 new template ID’s (TID)

l Used to convey exposure

characteristics and dose

generated by imaging devices

  • Therapeutic dose is reported in RT

Dose IOD

  • Records “radiation output”

information, not patient dose

TID 1002

Observer Context

TID Tx

Irradiation Event

X-Ray Data

TID Tx

Accumulated

X-Ray Dose Data

TID Tx

Projection X-Ray

Radiation Dose

TID 1002

Observer Context

TID 10003

Irradiation Event

X-Ray Data

TID 10002

Accumulated

X-Ray Dose Data

TID 10001

Projection X-Ray

Radiation Dose

Radiation Dose SR – Projection Radiography

l Mammography Dose SR was

developed in 2007

  • Allows templates for different

modalities or different modality

configurations to be included

NL Rel with
Parent
VT Concept Name VM Req
Type
Condition Value Set Constraint
1 CONTAINER EV (113702, DCM,

”Accumulated X-Ray Dose Data”)

1 M
2 > HAS
CONCEPT
MOD
CODE EV (113764, DCM,

“Acquisition Plane”

1 M $Plane

3 > CONTAINS TEXT EV (113780, DCM,

“Reference Point Definition”)

1 U
4 > CONTAINS CONTAINER EV (122505, DCM,

“Calibration”)

1 M
5 >> HAS
CONCEPT
MOD
CODE EV (113794, DCM,

“Dose Measurement Device”)

1-n M DCID (10010) Dose Measurement Devices

6 >> CONTAINS DATETIME EV (113723, DCM,

“Calibration Date”)

1 M
7 >> CONTAINS NUM EV (122322, DCM,

“Calibration Factor”)

1 M Units = EV (1, UCUM, “no units”)

8 >> CONTAINS NUM EV (113763, DCM,

“Calibration Uncertainty”)

1 M Units = EV (%, UCUM, “Percent”)

9 >> CONTAINS TEXT EV (113724, DCM,

“Calibration Responsible Party”)

1 M
10 > CONTAINS INCLUDE DTID (10004)

Accumulated Projection X-Ray Dose

1 MC XOR row 11, IFF TID (10001) Row 2 = (113704, DCM, “Projection X-Ray”)

11 > CONTAINS INCLUDE DTID (10005)

Accumulated Mammography X-Ray Dose

1 MC XOR row 10, IFF TID (10001) Row 2 = (P5-40010, SRT, “Mammography”)

10 > CONTAINS NUM EV (113722, DCM,

“Dose Area Product Total”)

1 M Units = EV (Gym2, UCUM, “Gym2”)

11 > CONTAINS NUM EV (113725, DCM,

“Dose (RP) Total”)

1 M Units = EV (Gy, UCUM, “Gy”)

12 > CONTAINS NUM EV (113726, DCM,

“Fluoro Dose Area Product Total”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event

Units = EV (Gym2, UCUM, “Gym2”)

13 > CONTAINS NUM EV (113728, DCM,

“Fluoro Dose (RP) Total”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event

Units = EV (Gy, UCUM, “Gy”)

14 > CONTAINS NUM EV (113730, DCM,

“Total Fluoro Time”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event.

Units = EV (s, UCUM, “s”)

15 > CONTAINS NUM EV (113727, DCM,

“Acquisition Dose Area Product Total”)

1 M Units = EV (Gym2, UCUM, “Gym2”)

16 > CONTAINS NUM EV (113729, DCM,

“Acquisition Dose (RP) Total”)

1 M Units = EV (Gy, UCUM, “Gy”)

17 > CONTAINS NUM EV (113731, DCM,

“Total Number of Radiographic Frames”)

1 U Units = EV (1, UCUM, “no units”)

TID 10002

ACCUMULATED X-RAY DOSE

This change (i.e. CP687) removed the

original projection radiography

accumulated dose information from

TID 10002

NL Rel with
Parent
VT Concept Name VM Req
Type
Condition Value Set Constraint
1 CONTAINER EV (113702, DCM,

”Accumulated X-Ray Dose Data”)

1 M
2 > HAS
CONCEPT
MOD
CODE EV (113764, DCM,

“Acquisition Plane”

1 M $Plane

3 > CONTAINS TEXT EV (113780, DCM,

“Reference Point Definition”)

1 U
4 > CONTAINS CONTAINER EV (122505, DCM,

“Calibration”)

1 M
5 >> HAS
CONCEPT
MOD
CODE EV (113794, DCM,

“Dose Measurement Device”)

1-n M DCID (10010) Dose Measurement Devices

6 >> CONTAINS DATETIME EV (113723, DCM,

“Calibration Date”)

1 M
7 >> CONTAINS NUM EV (122322, DCM,

“Calibration Factor”)

1 M Units = EV (1, UCUM, “no units”)

8 >> CONTAINS NUM EV (113763, DCM,

“Calibration Uncertainty”)

1 M Units = EV (%, UCUM, “Percent”)

9 >> CONTAINS TEXT EV (113724, DCM,

“Calibration Responsible Party”)

1 M
10 > CONTAINS INCLUDE DTID (10004)

Accumulated Projection X-Ray Dose

1 MC XOR row 11, IFF TID (10001) Row 2 = (113704, DCM, “Projection X-Ray”)

11 > CONTAINS INCLUDE DTID (10005)

Accumulated Mammography X-Ray Dose

1 MC XOR row 10, IFF TID (10001) Row 2 = (P5-40010, SRT, “Mammography”)

10 > CONTAINS NUM EV (113722, DCM,

“Dose Area Product Total”)

1 M Units = EV (Gym2, UCUM, “Gym2”)

11 > CONTAINS NUM EV (113725, DCM,

“Dose (RP) Total”)

1 M Units = EV (Gy, UCUM, “Gy”)

12 > CONTAINS NUM EV (113726, DCM,

“Fluoro Dose Area Product Total”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event

Units = EV (Gym2, UCUM, “Gym2”)

13 > CONTAINS NUM EV (113728, DCM,

“Fluoro Dose (RP) Total”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event

Units = EV (Gy, UCUM, “Gy”)

14 > CONTAINS NUM EV (113730, DCM,

“Total Fluoro Time”)

1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event.

Units = EV (s, UCUM, “s”)

15 > CONTAINS NUM EV (113727, DCM,

“Acquisition Dose Area Product Total”)

1 M Units = EV (Gym2, UCUM, “Gym2”)

16 > CONTAINS NUM EV (113729, DCM,

“Acquisition Dose (RP) Total”)

1 M Units = EV (Gy, UCUM, “Gy”)

17 > CONTAINS NUM EV (113731, DCM,

“Total Number of Radiographic Frames”)

1 U Units = EV (1, UCUM, “no units”)

Radiation Dose SR – Projection Radiography

l TID 10004

  • Accumulated dose information that

was in TID 10002 for projection

radiography systems

l TID 10005

  • Mammography Dose
    • Accumulated average glandular

dose and laterality

TID 1002

Observer Context

TID Tx

Irradiation Event

X-Ray Data

TID Tx

Accumulated

X-Ray Dose Data

TID Tx

Projection X-Ray

Radiation Dose

TID 1002

Observer Context

TID 10003

Irradiation Event

X-Ray Data

TID 10002

Accumulated

X-Ray Dose Data

TID 10001

Projection X-Ray

Radiation Dose

TID 10005

Accumulated

Mammography X-Ray

Dose

TID 10004

Accumulated

Projection X-Ray

Dose

Radiation Dose SR – TID 10003a

l Irradiation Event X-Ray Detector Data

  • 2008 AAPM TG116 and IEC 62494-1 exposure indicator values

VT Concept Name VM

Req

Type Condition Value Set Constraint

1 NUM EV (113845, DCM,

“Exposure Index”)

1 MC IF the value is displayable

to the X-ray system

operator.

Units = (1,UCUM, “no

units”)

2 NUM EV (113846, DCM,

“Target Exposure Index”)

1 MC IF the value is displayable

to the X-ray system

operator.

Units = (1,UCUM, “no

units”)

3 NUM EV (113847, DCM,

“Deviation Index”)

1 MC IF the value is displayable

to the X-ray system

operator.

Units = (1,UCUM, “no

units”)

4 INCLUDE DTID (1021) Device

Participant

1 U $DeviceProcedureRole =

EV (113942, DCM, “X-Ray

Reading Device”)

5 IMAGE EV (113795, DCM,

“Acquired Image”)

1-n MC IFF Image Object is

created for this irradiation

event

Reference to image

instances created

during each event

Radiation Dose SR – TID 10003b

l Irradiation Event X-Ray Source Data

  • kVp, mA, exposure time, focal spot, filter material etc.
  • Reference point location where dose is determined

Coding Scheme

Designator

(0008,0102)

Code Value

(0008,0100)

Code Meaning

(0008,0104)

DCM 113860 15cm from Isocenter toward Source

DCM 113861 30cm in Front of Image Input Surface

DCM 113862 1cm above Tabletop

DCM 113863 30cm above Tabletop

DCM 113864 15cm from Table Centerline

DCM 113865 Entrance exposure to a 4.2 cm breast

thickness

DCM 113941 In Detector Plane

CID 10025 Radiation Dose Reference Points

Radiation Dose SR – TID 10006

l Accumulated Cassette-Based Projection Radiography Dose

  • Count of the number of radiographic frames (i.e. images) recorded

for a study or the number of exposures

  • Whichever is “known” to be greater
  • Provides a method to determine if more exposures were produced than

images stored (retake analysis?)

  • Detector type specified

Coding Scheme

Designator

(0008,0102)

Code Value

(0008,0100)

Code Meaning

(0008,0104)

DCM 113948 Direct Detector

DCM 113949 Indirect Detector

DCM 113950 Storage Detector

DCM 113951 Film

CID 10030 Detector Types

Stores a signal which is

later transformed by a

reader into pixel values

(e.g. storage phosphor)

Radiation Dose SR – TID 10007

l Accumulated Integrated Projection Radiography Dose

  • Accumulated dose and dose area product

VT Concept Name VM Req

Type

Condition Value Set Constraint

1 NUM EV (113722, DCM, “Dose

Area Product Total”)

1 M Units = EV (Gy.m2, UCUM,

“Gy.m2”)

2 NUM EV (113725, DCM, “Dose

(RP) Total”)

1 M Units = EV (Gy, UCUM,

“Gy”)

3 NUM EV (113731, DCM, “Total

Number of Radiographic

Frames”)

1 U Units = EV (1, UCUM, “no

units”)

4 CODE EV (113780, DCM,

“Reference Point

Definition”)

1 MC IF Row 5 is not present. DCID (10025) Radiation

Dose Reference Points

5 TEXT EV (113780, DCM,

“Reference Point

Definition”)

1 MC IF Row 4 is not present.

Includes

reference point

Includes count of

exposures taken

(retake analysis?)

Radiation Dose SR – TID 10014 Scanning Length

l Added in 2009

l Defines detail relative to the scanning length for spiral

acquisitions due to the additional rotations at the beginning and

end of the scan

  • so-called “overranging”

Radiation Dose SR – TID 10015 Dose check values

l Records details related to the use of the NEMA Dose Check

Standard (NEMA XR-25-2010)

  • see http://www.aapm.org/pubs/CTProtocols/

14

VT Concept Name VM

Req

Type Condition

1 CONTAINER EV (113900, DCM, “Dose Check Alert

Details”)

1 MC IF the scanning device has implemented dose alerts

4 NUM EV (113903, DCM, “DLP Alert Value”) 1 MC IFF value of Row 2 is (R-0038D,SRT, “Yes”)

5 NUM EV (113904, DCM, “CTDIvol Alert

Value”)

1 MC IFF value of Row 3 is (R-0038D,SRT, “Yes”)

6 NUM EV(113905, DCM, “Accumulated DLP

Forward Estimate”)

1 MC IF Accumulated DLP Forward Estimate (Row 6)

exceeds DLP Alert Value (Row 4)

7 NUM EV (113906, DCM, “Accumulated

CTDIvol Forward Estimate”)

1 MC IF Accumulated CTDIvol Forward Estimate (Row 7)

exceeds CTDIvol Alert Value (Row 5)

8 TEXT EV (113907, DCM, “Reason for

Proceeding”)

1 UC IFF Accumulated DLP Forward Estimate (Row 6)

exceeds DLP Alert Value (Row 4) or Accumulated

CTDIvol Forward Estimate (Row 7) exceeds CTDIvol

Alert Value (Row 5)

  • • • • • •

Radiation Dose SR – Projection Radiography

Data

element tag

ID values

are the

same

Radiation Dose Information in Radiology Reports SR

l Approved August 2011

l TID 2006 Imaging Report with Conditional Radiation

exposure and Protection Information

  • Template used for general imaging reports
  • Specialization of TID 2000 Basic Diagnostic Imaging Report

l TID 2007 Imaging Procedure Description

l TID 2008 Radiation Exposure and Protection Information

NM RDSR Supplement 159

l Reporting of radiation dose due to administration of

radiopharmaceuticals, as in NM or PET imaging

  • Introduces 5 new templates

l Version out for public comment in April 2013

  • AAPM argued to not make ‘effective dose’ mandatory and SNMMI

commented to not include either ‘effective’ or ‘organ absorbed’

dose and if included it should be optional

  • Revisions underway responding to all comments

AAPM and DICOM Efforts

l WG28 Physics

  • Formed by DICOM Standards Committee (DSC) at RSNA 2011
  • Work Item to create Patient Radiation Dose Structured Report

approved by DSC at RSNA 2012