Donald Peck, PhD
Henry Ford Health System
Detroit, Michigan USA
Radiation Dose Structured Report
Current and Future
Radiation Dose SR
l Original developed in 2005 for
Projection Radiography
- Added 3 new template ID’s (TID)
l Used to convey exposure
characteristics and dose
generated by imaging devices
- Therapeutic dose is reported in RT
Dose IOD
- Records “radiation output”
information, not patient dose
TID 1002
Observer Context
TID Tx
Irradiation Event
X-Ray Data
TID Tx
Accumulated
X-Ray Dose Data
TID Tx
Projection X-Ray
Radiation Dose
TID 1002
Observer Context
TID 10003
Irradiation Event
X-Ray Data
TID 10002
Accumulated
X-Ray Dose Data
TID 10001
Projection X-Ray
Radiation Dose
Radiation Dose SR – Projection Radiography
l Mammography Dose SR was
developed in 2007
- Allows templates for different
modalities or different modality
configurations to be included
NL Rel with
Parent
VT Concept Name VM Req
Type
Condition Value Set Constraint
1 CONTAINER EV (113702, DCM,
”Accumulated X-Ray Dose Data”)
1 M
2 > HAS
CONCEPT
MOD
CODE EV (113764, DCM,
“Acquisition Plane”
1 M $Plane
3 > CONTAINS TEXT EV (113780, DCM,
“Reference Point Definition”)
1 U
4 > CONTAINS CONTAINER EV (122505, DCM,
“Calibration”)
1 M
5 >> HAS
CONCEPT
MOD
CODE EV (113794, DCM,
“Dose Measurement Device”)
1-n M DCID (10010) Dose Measurement Devices
6 >> CONTAINS DATETIME EV (113723, DCM,
“Calibration Date”)
1 M
7 >> CONTAINS NUM EV (122322, DCM,
“Calibration Factor”)
1 M Units = EV (1, UCUM, “no units”)
8 >> CONTAINS NUM EV (113763, DCM,
“Calibration Uncertainty”)
1 M Units = EV (%, UCUM, “Percent”)
9 >> CONTAINS TEXT EV (113724, DCM,
“Calibration Responsible Party”)
1 M
10 > CONTAINS INCLUDE DTID (10004)
Accumulated Projection X-Ray Dose
1 MC XOR row 11, IFF TID (10001) Row 2 = (113704, DCM, “Projection X-Ray”)
11 > CONTAINS INCLUDE DTID (10005)
Accumulated Mammography X-Ray Dose
1 MC XOR row 10, IFF TID (10001) Row 2 = (P5-40010, SRT, “Mammography”)
10 > CONTAINS NUM EV (113722, DCM,
“Dose Area Product Total”)
1 M Units = EV (Gym2, UCUM, “Gym2”)
11 > CONTAINS NUM EV (113725, DCM,
“Dose (RP) Total”)
1 M Units = EV (Gy, UCUM, “Gy”)
12 > CONTAINS NUM EV (113726, DCM,
“Fluoro Dose Area Product Total”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event
Units = EV (Gym2, UCUM, “Gym2”)
13 > CONTAINS NUM EV (113728, DCM,
“Fluoro Dose (RP) Total”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event
Units = EV (Gy, UCUM, “Gy”)
14 > CONTAINS NUM EV (113730, DCM,
“Total Fluoro Time”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event.
Units = EV (s, UCUM, “s”)
15 > CONTAINS NUM EV (113727, DCM,
“Acquisition Dose Area Product Total”)
1 M Units = EV (Gym2, UCUM, “Gym2”)
16 > CONTAINS NUM EV (113729, DCM,
“Acquisition Dose (RP) Total”)
1 M Units = EV (Gy, UCUM, “Gy”)
17 > CONTAINS NUM EV (113731, DCM,
“Total Number of Radiographic Frames”)
1 U Units = EV (1, UCUM, “no units”)
TID 10002
ACCUMULATED X-RAY DOSE
This change (i.e. CP687) removed the
original projection radiography
accumulated dose information from
TID 10002
NL Rel with
Parent
VT Concept Name VM Req
Type
Condition Value Set Constraint
1 CONTAINER EV (113702, DCM,
”Accumulated X-Ray Dose Data”)
1 M
2 > HAS
CONCEPT
MOD
CODE EV (113764, DCM,
“Acquisition Plane”
1 M $Plane
3 > CONTAINS TEXT EV (113780, DCM,
“Reference Point Definition”)
1 U
4 > CONTAINS CONTAINER EV (122505, DCM,
“Calibration”)
1 M
5 >> HAS
CONCEPT
MOD
CODE EV (113794, DCM,
“Dose Measurement Device”)
1-n M DCID (10010) Dose Measurement Devices
6 >> CONTAINS DATETIME EV (113723, DCM,
“Calibration Date”)
1 M
7 >> CONTAINS NUM EV (122322, DCM,
“Calibration Factor”)
1 M Units = EV (1, UCUM, “no units”)
8 >> CONTAINS NUM EV (113763, DCM,
“Calibration Uncertainty”)
1 M Units = EV (%, UCUM, “Percent”)
9 >> CONTAINS TEXT EV (113724, DCM,
“Calibration Responsible Party”)
1 M
10 > CONTAINS INCLUDE DTID (10004)
Accumulated Projection X-Ray Dose
1 MC XOR row 11, IFF TID (10001) Row 2 = (113704, DCM, “Projection X-Ray”)
11 > CONTAINS INCLUDE DTID (10005)
Accumulated Mammography X-Ray Dose
1 MC XOR row 10, IFF TID (10001) Row 2 = (P5-40010, SRT, “Mammography”)
10 > CONTAINS NUM EV (113722, DCM,
“Dose Area Product Total”)
1 M Units = EV (Gym2, UCUM, “Gym2”)
11 > CONTAINS NUM EV (113725, DCM,
“Dose (RP) Total”)
1 M Units = EV (Gy, UCUM, “Gy”)
12 > CONTAINS NUM EV (113726, DCM,
“Fluoro Dose Area Product Total”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event
Units = EV (Gym2, UCUM, “Gym2”)
13 > CONTAINS NUM EV (113728, DCM,
“Fluoro Dose (RP) Total”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event
Units = EV (Gy, UCUM, “Gy”)
14 > CONTAINS NUM EV (113730, DCM,
“Total Fluoro Time”)
1 MC IFF TID(10003) Row 3 value = (P5-06000, SRT, “Fluoroscopy”) for at least one irradiation event.
Units = EV (s, UCUM, “s”)
15 > CONTAINS NUM EV (113727, DCM,
“Acquisition Dose Area Product Total”)
1 M Units = EV (Gym2, UCUM, “Gym2”)
16 > CONTAINS NUM EV (113729, DCM,
“Acquisition Dose (RP) Total”)
1 M Units = EV (Gy, UCUM, “Gy”)
17 > CONTAINS NUM EV (113731, DCM,
“Total Number of Radiographic Frames”)
1 U Units = EV (1, UCUM, “no units”)
Radiation Dose SR – Projection Radiography
l TID 10004
- Accumulated dose information that
was in TID 10002 for projection
radiography systems
l TID 10005
- Mammography Dose
- Accumulated average glandular
dose and laterality
TID 1002
Observer Context
TID Tx
Irradiation Event
X-Ray Data
TID Tx
Accumulated
X-Ray Dose Data
TID Tx
Projection X-Ray
Radiation Dose
TID 1002
Observer Context
TID 10003
Irradiation Event
X-Ray Data
TID 10002
Accumulated
X-Ray Dose Data
TID 10001
Projection X-Ray
Radiation Dose
TID 10005
Accumulated
Mammography X-Ray
Dose
TID 10004
Accumulated
Projection X-Ray
Dose
Radiation Dose SR – TID 10003a
l Irradiation Event X-Ray Detector Data
- 2008 AAPM TG116 and IEC 62494-1 exposure indicator values
VT Concept Name VM
Req
Type Condition Value Set Constraint
1 NUM EV (113845, DCM,
“Exposure Index”)
1 MC IF the value is displayable
to the X-ray system
operator.
Units = (1,UCUM, “no
units”)
2 NUM EV (113846, DCM,
“Target Exposure Index”)
1 MC IF the value is displayable
to the X-ray system
operator.
Units = (1,UCUM, “no
units”)
3 NUM EV (113847, DCM,
“Deviation Index”)
1 MC IF the value is displayable
to the X-ray system
operator.
Units = (1,UCUM, “no
units”)
4 INCLUDE DTID (1021) Device
Participant
1 U $DeviceProcedureRole =
EV (113942, DCM, “X-Ray
Reading Device”)
5 IMAGE EV (113795, DCM,
“Acquired Image”)
1-n MC IFF Image Object is
created for this irradiation
event
Reference to image
instances created
during each event
Radiation Dose SR – TID 10003b
l Irradiation Event X-Ray Source Data
- kVp, mA, exposure time, focal spot, filter material etc.
- Reference point location where dose is determined
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
DCM 113860 15cm from Isocenter toward Source
DCM 113861 30cm in Front of Image Input Surface
DCM 113862 1cm above Tabletop
DCM 113863 30cm above Tabletop
DCM 113864 15cm from Table Centerline
DCM 113865 Entrance exposure to a 4.2 cm breast
thickness
DCM 113941 In Detector Plane
CID 10025 Radiation Dose Reference Points
Radiation Dose SR – TID 10006
l Accumulated Cassette-Based Projection Radiography Dose
- Count of the number of radiographic frames (i.e. images) recorded
for a study or the number of exposures
- Whichever is “known” to be greater
- Provides a method to determine if more exposures were produced than
images stored (retake analysis?)
Coding Scheme
Designator
(0008,0102)
Code Value
(0008,0100)
Code Meaning
(0008,0104)
DCM 113948 Direct Detector
DCM 113949 Indirect Detector
DCM 113950 Storage Detector
DCM 113951 Film
CID 10030 Detector Types
Stores a signal which is
later transformed by a
reader into pixel values
(e.g. storage phosphor)
Radiation Dose SR – TID 10007
l Accumulated Integrated Projection Radiography Dose
- Accumulated dose and dose area product
VT Concept Name VM Req
Type
Condition Value Set Constraint
1 NUM EV (113722, DCM, “Dose
Area Product Total”)
1 M Units = EV (Gy.m2, UCUM,
“Gy.m2”)
2 NUM EV (113725, DCM, “Dose
(RP) Total”)
1 M Units = EV (Gy, UCUM,
“Gy”)
3 NUM EV (113731, DCM, “Total
Number of Radiographic
Frames”)
1 U Units = EV (1, UCUM, “no
units”)
4 CODE EV (113780, DCM,
“Reference Point
Definition”)
1 MC IF Row 5 is not present. DCID (10025) Radiation
Dose Reference Points
5 TEXT EV (113780, DCM,
“Reference Point
Definition”)
1 MC IF Row 4 is not present.
Includes
reference point
Includes count of
exposures taken
(retake analysis?)
Radiation Dose SR – TID 10014 Scanning Length
l Added in 2009
l Defines detail relative to the scanning length for spiral
acquisitions due to the additional rotations at the beginning and
end of the scan
Radiation Dose SR – TID 10015 Dose check values
l Records details related to the use of the NEMA Dose Check
Standard (NEMA XR-25-2010)
- see http://www.aapm.org/pubs/CTProtocols/
14
VT Concept Name VM
Req
Type Condition
1 CONTAINER EV (113900, DCM, “Dose Check Alert
Details”)
1 MC IF the scanning device has implemented dose alerts
4 NUM EV (113903, DCM, “DLP Alert Value”) 1 MC IFF value of Row 2 is (R-0038D,SRT, “Yes”)
5 NUM EV (113904, DCM, “CTDIvol Alert
Value”)
1 MC IFF value of Row 3 is (R-0038D,SRT, “Yes”)
6 NUM EV(113905, DCM, “Accumulated DLP
Forward Estimate”)
1 MC IF Accumulated DLP Forward Estimate (Row 6)
exceeds DLP Alert Value (Row 4)
7 NUM EV (113906, DCM, “Accumulated
CTDIvol Forward Estimate”)
1 MC IF Accumulated CTDIvol Forward Estimate (Row 7)
exceeds CTDIvol Alert Value (Row 5)
8 TEXT EV (113907, DCM, “Reason for
Proceeding”)
1 UC IFF Accumulated DLP Forward Estimate (Row 6)
exceeds DLP Alert Value (Row 4) or Accumulated
CTDIvol Forward Estimate (Row 7) exceeds CTDIvol
Alert Value (Row 5)
Radiation Dose SR – Projection Radiography
Data
element tag
ID values
are the
same
Radiation Dose Information in Radiology Reports SR
l Approved August 2011
l TID 2006 Imaging Report with Conditional Radiation
exposure and Protection Information
- Template used for general imaging reports
- Specialization of TID 2000 Basic Diagnostic Imaging Report
l TID 2007 Imaging Procedure Description
l TID 2008 Radiation Exposure and Protection Information
NM RDSR Supplement 159
l Reporting of radiation dose due to administration of
radiopharmaceuticals, as in NM or PET imaging
- Introduces 5 new templates
l Version out for public comment in April 2013
- AAPM argued to not make ‘effective dose’ mandatory and SNMMI
commented to not include either ‘effective’ or ‘organ absorbed’
dose and if included it should be optional
- Revisions underway responding to all comments
AAPM and DICOM Efforts
l WG28 Physics
- Formed by DICOM Standards Committee (DSC) at RSNA 2011
- Work Item to create Patient Radiation Dose Structured Report
approved by DSC at RSNA 2012