EVALUATION OF HAMSTRING FLEXIBILITY BY USING TWO ..., Exercises of Human Biology

Hamstring flexibility was evaluated by two different methods of measurement: a) a Myrin goni- ometer and b) sit and reach test.

Typology: Exercises

2022/2023

Uploaded on 03/01/2023

ekani
ekani 🇺🇸

4.7

(26)

265 documents

1 / 7

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
Bakirtzoglou, P., Ioannou, P. & Bakirtzoglou, F.: EVALUATION OF HAMSTRING... SportLogia 6 (2010) 2: 28-32
28
EVALUATION OF HAMSTRING FLEXIBILITY BY USING
TWO DIFFERENT MEASURING INSTRUMENTS
Bakirtzoglou Panteleimon1,Ioannou Panagiotis2& Bakirtzoglou Fotis3
1Organisation for Vocational Education and Training in Greece, Athens, Greece
2Faculty of Physical Education and Sports Science, Thessaloniki, Greece
3General Hospital of Thessaloniki "Agios Dimitrios", Thesaloniki, Greece
SHORT SCIENTIFIC ARTICLE
DOI: 10.5550/sgia.1002028 COBISS.BH-ID 1846808 UDC: 616.728.3:796.012.23
SUMMARY
The purpose of the present study was to investigate the effect of two different methods of
measurement for hamstring flexibility. Forty male students athletes with mean age 23.45±0.44
years and forty non-athletes students with a mean age 23.08±0.98 years participated in this study.
Hamstring flexibility was evaluated by two different methods of measurement: a) a Myrin goni-
ometer and b) sit and reach test. Statistical analysis included the use of Independent Samples T-
test while significance was set at p<0.01. The results indicated that athletes students scored better
than non-athletes students only when hip joint’s mobility was measured with a Myrin goniometer.
In conclusion the evaluation of joint's mobility should be done by using a method of measure-
ment that would isolate the articulation of measurement from the interjection of other joints or
muscular teams something that is achieved by the use of Myrin goniometer than the use of Sit
and Reach test.
Key words: hamstrings, Myrin goniometer, sit and reach test.
INTRODUCTION
Sports performance is the result of the
interplay of various mental and physical
factors. One of the main factors to improve
and maximize athletic performance is the
development of physical abilities. The capabi-
lities of fitness are strength, speed, endurance
and flexibility (American College of Sports
Medicine, 1995).
Flexibility has been defined in many ways.
It is used the term joint mobility indicates the
degree of bending the normal range or scope
of a joint or series of joints (Moller & Oberg,
1984).
Several sit-and-reach tests (SRs) are com-
monly used in health-related and physical-
fitness test batteries to evaluate the hamstring
and lower back flexibility (Jackson & Baker,
1986; Hoeger et al, 1990; Hui and Yuen,
2000). Such field measures are only moderate
indicators of hamstring extensibility. However,
the SRs are frequently used to evaluate the
hamstring muscle extensibility because the
procedures are simple, easy to administer, re-
quire minimal skills training and are particu-
larly useful in large scale extensibility evalua-
tion in the field setting. The classical SR (Wells
& Dillon, 1952) was originally selected as a
part of the American Alliance for Health,
Physical Education, Recreation & Dance
(AAHPERD) health-related and physical- fit-
ness protocol, and is often included in stand-
ard fitness tests (e.g., EUROFIT: personal
fitness tests, President’s Challenge) and health
related fitness programs (Institute for Aerobic
Research, 1988) and Α.A.H.P.E.R.D.
(AAHPERD, 1984).
pf3
pf4
pf5

Partial preview of the text

Download EVALUATION OF HAMSTRING FLEXIBILITY BY USING TWO ... and more Exercises Human Biology in PDF only on Docsity!

EVALUATION OF HAMSTRING FLEXIBILITY BY USING

TWO DIFFERENT MEASURING INSTRUMENTS

Bakirtzoglou Panteleimon^1 , Ioannou Panagiotis^2 & Bakirtzoglou Fotis^3 (^1) Organisation for Vocational Education and Training in Greece, Athens, Greece (^2) Faculty of Physical Education and Sports Science, Thessaloniki, Greece (^3) General Hospital of Thessaloniki "Agios Dimitrios", Thesaloniki, Greece SHORT SCIENTIFIC ARTICLE DOI: 10.5550/sgia.1002028 COBISS.BH-ID 1846808 UDC: 616.728.3:796.012.

SUMMARY

The purpose of the present study was to investigate the effect of two different methods of measurement for hamstring flexibility. Forty male students athletes with mean age 23.45±0. years and forty non-athletes students with a mean age 23.08±0.98 years participated in this study. Hamstring flexibility was evaluated by two different methods of measurement: a) a Myrin goni- ometer and b) sit and reach test. Statistical analysis included the use of Independent Samples T- test while significance was set at p<0.01. The results indicated that athletes students scored better than non-athletes students only when hip joint’s mobility was measured with a Myrin goniometer. In conclusion the evaluation of joint's mobility should be done by using a method of measure- ment that would isolate the articulation of measurement from the interjection of other joints or muscular teams something that is achieved by the use of Myrin goniometer than the use of Sit and Reach test. Key words: hamstrings, Myrin goniometer, sit and reach test.

INTRODUCTION

Sports performance is the result of the interplay of various mental and physical factors. One of the main factors to improve and maximize athletic performance is the development of physical abilities. The capabi- lities of fitness are strength, speed, endurance and flexibility (American College of Sports Medicine, 1995). Flexibility has been defined in many ways. It is used the term joint mobility indicates the degree of bending the normal range or scope of a joint or series of joints (Moller & Oberg, 1984). Several sit-and-reach tests (SRs) are com- monly used in health-related and physical- fitness test batteries to evaluate the hamstring and lower back flexibility (Jackson & Baker, 1986; Hoeger et al, 1990; Hui and Yuen, 2000). Such field measures are only moderate indicators of hamstring extensibility. However, the SRs are frequently used to evaluate the hamstring muscle extensibility because the procedures are simple, easy to administer, re- quire minimal skills training and are particu- larly useful in large scale extensibility evalua- tion in the field setting. The classical SR (Wells & Dillon, 1952) was originally selected as a part of the American Alliance for Health, Physical Education, Recreation & Dance (AAHPERD) health-related and physical- fit- ness protocol, and is often included in stand- ard fitness tests (e.g., EUROFIT: personal fitness tests, President’s Challenge) and health related fitness programs (Institute for Aerobic Research, 1988) and Α.A.H.P.E.R.D. (AAHPERD, 1984).

During measuring joint’s mobility, how- ever, a joint should be used as instruments to limit interference with other muscle groups or joints in the final measurement result. According to Harris (1969), the joint must be isolated at the time of measurement. Myrin goniometer isolates the joint from the interference of other joints without affecting the final outcome of measuring the length of other muscle groups (Leighton, 1955). Unlike measurement with the Sit and Reach test may lead to different results, because in anatomy and motion analysis of the test, participating in the final outcome of the measurement back muscles, hamstrings and the triceps surae muscle complex (Kendall et al, 1971). Most research has not dealt with the issue of the effect of the measure instrument to the final result of the measurement of a joint mobility. The purpose of the present study was to examine whether the use of two different instruments for the evaluation of knee joint’s mobility may affect the final result of measurement between untrained and trained individuals.

METHOD

Participants

Forty student-athletes and forty students (non athletes) participated in this study. Stu- dent-athletes had an average age of 23.45 ± 0.44 years, height 1.89 ± 0.69cm and weight 84.56 ± 12.43kg. The relevant anthropometric characteristic of students (non-athletes) was: age, years 23.08 ± 0.98, height 1.86 ± 0.97cm and weight 91 ± 13.46kg. Both student- athletes and students (non-athletes) do not have a musculoskeletal problem before the time of measurement.

Procedure

Testing took place in the exercise room at a private gym. Before testing, all participants performed a three minute warm up and static stretch routine, emphasizing the lower body. Immediately after the stretching, the flexibility tests were performed in a counterbalanced design. All tests were assessed on the same day for each student. The participants were allowed to rest for 20 minutes between tests. One physiotherapist was responsible for each test. All measures were performed on the same day, and all tests were conducted with the participants wearing their shoes. For the evaluation of knee joint’s mobility have been used two different instruments: the Myrin goniometer (Leighton, 1955) and Sit and Reach test (Wells & Dillon, 1952). The measurement of hip flexion with knee straight with Myrin goniometer was tested in a medical bed. The initial and final position of each movement passively measured starting from the anatomical neutral point O, as determined by the American Academy of Orthopedic Surgeons (1965). Testing room temperature was kept at 25 °C.

Measures

Goniometric Measurement The Myrin goniometer (Lic Rehab. 17183 Solna, Sweden) is a goniometer vertical and horizontal measurements based on Leighton flexometer (1955). It consists of a circular range (0-180th) degrees turntable and two indices. One indicator is mounted in the center of the disc controlled by gravity for vertical measurements and the other is an indicator of orientation for horizontal measu- rements. The measurement was performed using American Academy of Orthopedic Surgeons (1965) procedures. A manual goniometer was used to measure flexibility as the range of motion in the hip joint and related musculature, while the participant lay supine on a firm, level examining table. The examiner raised the tested leg slowly and evenly, with the knee fully extended, avoiding abduction and rotation, until tightness or pain restricted the movement. The upward motion of the straight leg was measured to the nearest degree from the zero starting position. The hamstring flexibility score was determined as the mean of two measurements of the left and right legs. Ekstrand et al. (1982) demonstrated the reliability of measuring joint range of motion as a measure of hamstring flexibility.

passive straight-leg raise on each leg, (Ekstrand, Wiktorsson, Oberg, & Gillquist, 1982; Jackson & Baker, 1986; Jackson & Langford, 1989; Minkler & Patterson, 1994). However, comparing these two instruments Kendall et al (1971) report that Sit and Reach test does not isolate the joint at the time of measurement and the final result would probably be due to physiological or limited length of the muscles of the knee, back and triceps surae muscle complex. Unlike the Myrin goniometer isolate the joint measurement, so the final result of measurement is the length of the extensor muscles (Moller & Oberg, 1984; Kippers & Parker, 1987). Most research has not dealt with the issue of the effect of the measure instrument to the final result of the measurement of a joint mobility. In this study used two different instruments in order to determine whether the use of two different instruments for evaluation hamstring flexibility will affect the final result of measurement. In this study, both students-athletes and students (non-athletes) had no statistically significant differences when used as an instrument for hamstring flexibility the Sit and Reach test. These results did not support the findings of Jackson and Baker (1986) and Chung and Yuen (1999) indicating that athletes have better statistical performance than non-athlete. Using the Sit and Reach test except for knee joint, back muscles and triceps surae muscle complex are involved (Kendall et al, 1971). Therefore the same performance between the two groups may be due to the physiological or limited length of back muscles, knee and triceps surae muscle complex. Rather, when used as measuring instrument the Myrin goniometer differences between the two groups was statistically significant in favor of student-athletes. This result is in agreement with Kendall et al (1971) and Jackson & Langford (1989) indicating that athletes perform better than untrained subjects in evaluation of flexibility with the Myrin goniometer. TABLE 2. Performance of student (non-athletes) and students-athletes in the evaluation of hamstring flexibility with the Myrin goniometer and the Sit and Reach test. The values are averages ± standard deviation. Students(non-athletes) Students-athletes Goniometer Myrin (o) 82 .00 ± 3.14 88 .00±2. Sit and reach test(cm) 21.98 ± 2.17 22.40 ± 1. p <.

CONCLUSION

Significant differences were only observed by the use of the Myrin goniometer for the evaluation of hamstring flexibility, in relation to Sit and Reach test, because it isolates the joint. In this study, knees joint mobility of trained and untrained individuals evaluated by the same score when used as a measure instrument the Sit and Reach test but this is likely due to the interference of other muscle groups in the measurement result. In contrast, trained student had higher results in the measurement of knees joint mobility from untrained student when using the Myrin goniometer, because the goniometer isolates the joint measurement, and the final result due to the length of the extensor muscles without involving other muscle groups or joints.

REFERENCE:

A.A.H.P.E.R.D. (1984). Technical manual for the health related physical fitness test. Washington, DC: Author. American Academy of Orthopedic Surgeons. (1965). Joint Motion: Method of measuring and recording. Chicago, IL: Park Ridge American College of Sports Medicine. (1995). Guidelines for exercise testing and prescription. Philadelphia: Lea and Febiger. Baltaci, G.N.UN, Tunay, V.A., Besler, S. & Gerceker, S. (2003). Comparison of three

different sit and reach tests for measure- ment of hamstring flexibility in female uni- versity students. Br. J. Sports Med. 37(1) , pp. 59 - 61. Chung, P.K. & Yuen, C.K. (1999). Criterion- related validity of sit-and-reach tests in uni- versity men in Hong Kong: Perceptual and Motor Skills. 88(1) , pp. 304 - 316. Ekstrand, J., Wiktorsson, M., Oberg, B, & Gillquist, J. (1982). Lower extremity goni- ometric measurements: A study to deter- mine their reliability. Arch. Phys. Med. Re- habil. 63(4) , pp. 171 - 175. Harris, M.L. (1969). A factor analytic study of flexibility. Res. Q. Exerc. Sport. 40 , pp. 62 - 70. Hoeger, W.W.K., Hopkins, D.R., Button, S. & Palmer, T.A. (1990). Comparing the sit and reach with the modified sit and reach in measuring flexibility in adolescents. Pediatric Exercise Science 2 , pp. 156 - 162. Hui, S.S.C. & Yuen, P.Y. (2000). Validity of the modified back-saver sit-and-reach test: a comparison with others protocols. Med. Sci. Sports Exerc. 32(9) , pp. 1655 - 1659. Institute for Aerobics Research (1988). The Fitnessgram. Dallas: Author. Jackson, A.W. & Baker, A.A. (1986). The rela- tionship of the sit and reach test to crite- rion measures of hamstring and back flexi- bility in young females. Res. Q. Exerc. Sport. 57 , pp. 183 - 186. Jackson, A.W. & Langford, N.J. (1989). The criterion-related validity of the sit and reach test: Replication and extension of previous findings. Res. Q. Exerc. Sport. 60(4) , pp. 384 -

Kendall, H.O., Kendall, F.P. & Wadsworth, G.E. (1971). Muscles: Testing and Function. (2nd ed.). Baltimore. MD: Williams and Wil- kins. Kippers, V. & Parker, A.W. (1987). Toe touch test. A measure of its validity. Physical Ther- apy, 67(11) , pp. 1680 - 84. Leighton, J.R. (1955). Instrument and tech- nique for measurement of range of joint motion. Archives Physical Medicine Rehabilita- tion, 36 , pp. 571 - 78. Liemohn, W., Sharpe, G.L. & Wasserman, J.F. (1994). Criterion related validity of the sit- and- reach test. J. Strength Cond Res. 8(2) , pp. 91 - 94. Minkler, S. & Patterson, P. (1994). The validity of the modified sit-and-reach test in col- lege-age students. Research Quarterly for Exer- cise and Sport 65 , pp. 189 - 192. Moller, B. & Oberg, B. (1984). Athletic train- ing and flexibility. A study on range of mo- tion in the lower extremity. Thesis, Linkoping. Simoneau, G.G. (1998). The impact of various anthropometric and flexibility measure- ments on the Sit-and-Reach test. J. Strength Cond Res. 12(4) , pp. 232 - 237. Wells, K.F. & Dillon, E.K. (1952). The sit and reach: A test of back and leg flexibility. Res. Q. for Exerc. Sport. 23 , pp. 115 - 118. Received: October, 20 th^2010 Accepted: December, 13 th^2010 Correspodence to: Panteleimon Bakirtzoglou Ph.D Organisation for Vocational Education and Training Ethikis Antistaseos 41 - OEEK TK.142 34 Athens-Nea Ionia Phone: +30 69 75 90 96 31 E-mail: [email protected]

sjedi-i-dohvati test ne izoluje zlob tokom mjerenja i da konačan rezultat vjerovatno zavisiti od ograničenja uslovljenih dužinom mišića leđa, koljena i triceps surae. Za razliku od toga Myrin goniometar izoluje zglob, tako da na konačan rezultat mjerenja utiče samo dužina mišića ekstenzora (Moller & Oberg, 1984; Kippers & Parker, 1987). U ovom istraživanju i studenti (sportisti) i student (nesportisti) nisu pokazali statistički značajnu razliku kada je kao instrument za procenu fleksibilnosti koljena korišćen sjedi-i-dohvati test. Ovi rezultati nisu potvrdili zaključke Jackson i Baker (1896) niti Chung i Yuen (1999) koji su pokazali da sportisti, statistički gledano, imaju bolje rezultate od nesportista. Koristeći sjedi-i-dohvati test, osim zgloba koljena, u pokret su takođe uključeni mišići leđa i triceps surae (Kendall et al, 1971). Stoga, razlika u mjerenju između dvije grupe je uslovljena uticajem mišića leđa, koljena i triceps surae na mjerenje. Drugim riječima, kada je korišćen Myrin goniometar razlike između dvije grupe su bile statistički značajne u korist studenata sportista. Ovaj rezultat je u saglasnosti sa Kendall et al (1971) i Jackson & Langford (1989) koji pokazuje da sportisti imaju bolje rezultate od netreniranih prilikom procjene fleksibilnosti sa Myrin goniometrom. Značajne razlike primjećene su samo pri korišćenju Myrin goniometra, u odnosu na sjedi-i-dohvati test. Razlog tome je što prvi test izoluje zglob. U ovom istraživanju fleksibilnost koljenog zgloba treniranih i netreniranih osoba procijenjena je sa potpuno istim rezultatom kada je kao merni instrument korišten sjedi-i-dohvati test, a uzrok tome je uticaj drugih mišićnih grupa na rezultate mjerenja. Ključne riječi: koljena, Myrin goniometar, sjedi-i-dohvati test.