Michelson Lab Report, Study Guides, Projects, Research of Physics

Physics Lab report, on Michelson Interferometer

Typology: Study Guides, Projects, Research

2017/2018

Uploaded on 12/12/2018

fardinanik
fardinanik 🇨🇳

5

(1)

1 document

1 / 10

Toggle sidebar

This page cannot be seen from the preview

Don't miss anything!

bg1
1/10
PHY 4264L Michelson Interferometer OPTICS LAB
LAB #6 The Michelson Interferometer
OBJECTIVE
To study the capabilities and uses of the Michelson interferometer; to use the interferometer to measure
wavelength of the HeNe laser and to measure the refractive index of air and invesitgate its dependence on
pressure.
MATERIALS
Pasco Optical Bench Vacuum cell with hand vacuum pump
Pasco Interferometer Beam Expander Lens with component holder
Movable mirror Pasco Component Holders
Beam splitter Adjustable Fixed Mirror
HeNe Laser MS Excel for Plotting
INTRODUCTION
INTERFERENCE THEORY
A beam of light can be modelled as a wave of oscillating electric and magnetic fields. When two beams of
light meet in space, these fields add according to the principle of superposition. At each point in space,
the electric and magnetic fields are determined as the vector sum of the fields of the separate beams.
If the two beams of light originate from separate sources, there is generally no fixed relationship between
the electromagnetic oscillations in the beams. If two such light beams meet, at any instant in time there
will be points in space where the fields add to produce a maximum field strength.
However, the oscillations of visible light are much faster than the human eye can apprehend. Since there
is no fixed relationship between the oscillations, a point at which there is a maximum at one instant
may have a minimum at the next instant. The human eye averages these results and perceives a uniform
intensity of light. However, if the two beams of light originate from the same source, there is generally
some degree of correlation between the frequency and phase of the oscillations of the two beams. At one
point in space the light from the beams may be continually in phase. In this case, the combined field will
always be a maximum and a bright spot will be seen. At another point the light from the two beams may
be continually out of phase and a minima, or dark spot, will be seen.
Thomas Young was one of the first to design a method for producing such an interference pattern. He
allowed a single, narrow beam of light to fall on two narrow, closely spaced slits. Opposite the slits he
placed a viewing screen. Where the light from the two slits struck the screen, a regular pattern of dark
and bright bands became visible. When first performed, Youngs experiment offered important evidence for
the wave nature of light.
Youngs slits function as a simple interferometer. If the spacing between the slits is known, the spacing of
the maxima and minima can be used to determine the wavelength of the light. Conversely, if the wavelength
of the light is known, the spacing of the slits could be determined from the interference patterns.
pf3
pf4
pf5
pf8
pf9
pfa

Partial preview of the text

Download Michelson Lab Report and more Study Guides, Projects, Research Physics in PDF only on Docsity!

LAB #6 – The Michelson Interferometer

OBJECTIVE

To study the capabilities and uses of the Michelson interferometer; to use the interferometer to measure wavelength of the HeNe laser and to measure the refractive index of air and invesitgate its dependence on pressure.

MATERIALS

◦ Pasco Optical Bench ◦ Vacuum cell with hand vacuum pump ◦ Pasco Interferometer ◦ Beam Expander Lens with component holder ◦ Movable mirror ◦ Pasco Component Holders ◦ Beam splitter ◦ Adjustable Fixed Mirror ◦ HeNe Laser ◦ MS Excel for Plotting

INTRODUCTION

INTERFERENCE THEORY

A beam of light can be modelled as a wave of oscillating electric and magnetic fields. When two beams of light meet in space, these fields add according to the principle of superposition. At each point in space, the electric and magnetic fields are determined as the vector sum of the fields of the separate beams.

If the two beams of light originate from separate sources, there is generally no fixed relationship between the electromagnetic oscillations in the beams. If two such light beams meet, at any instant in time there will be points in space where the fields add to produce a maximum field strength.

However, the oscillations of visible light are much faster than the human eye can apprehend. Since there is no fixed relationship between the oscillations, a point at which there is a maximum at one instant may have a minimum at the next instant. The human eye averages these results and perceives a uniform intensity of light. However, if the two beams of light originate from the same source, there is generally some degree of correlation between the frequency and phase of the oscillations of the two beams. At one point in space the light from the beams may be continually in phase. In this case, the combined field will always be a maximum and a bright spot will be seen. At another point the light from the two beams may be continually out of phase and a minima, or dark spot, will be seen.

Thomas Young was one of the first to design a method for producing such an interference pattern. He allowed a single, narrow beam of light to fall on two narrow, closely spaced slits. Opposite the slits he placed a viewing screen. Where the light from the two slits struck the screen, a regular pattern of dark and bright bands became visible. When first performed, Youngs experiment offered important evidence for the wave nature of light.

Youngs slits function as a simple interferometer. If the spacing between the slits is known, the spacing of the maxima and minima can be used to determine the wavelength of the light. Conversely, if the wavelength of the light is known, the spacing of the slits could be determined from the interference patterns.

THE MICHELSON INTERFEROMETER

In 1881, some 78 years after Young introduced his two-slit experiment, A.A. Michelson designed and built an interferometer using a similar principle. Originally Michelson designed his interferometer as a method to test for the existence of the ether, a hypothesized medium in which light could propagate. Due in part to his efforts, the ether is no longer considered a viable hypothesis. Michelsons interferometer has become a widely used instrument for measuring the wavelength of light, and for using the wavelength of a known light source to measure extremely small distances.

Figure 1 shows a diagram of a Michelson interferometer. A beam of light from the laser source strikes the beam-splitter. The beam-splitter is designed to reflect 50% of the incident light and transmit the other 50%. The incident beam therefore splits into two beams; one beam is reflected toward mirror M 1 , the other is transmitted toward mirror M 2. M 1 and M 2 reflect the beams back toward the beam-splitter. Half the light from M 1 is transmitted through the beam-splitter to the viewing screen and half the light from M 2 is reflected by the beam-splitter to the viewing screen.

Figure 1: Michelson Interferometer

In this way the original beam of light splits, and portions of the resulting beams are brought back together. The beams are from the same source and their phases highly correlate. The HeNe laser beam we use makes a small spot, so the interference is hard to see. To make it bigger we insert a lens between the laser and the beam splitter. When a lens is placed between the laser source and the beam-splitter, the light ray spreads out. An interference pattern of dark and bright rings, or fringes, is seen on the viewing screen, as shown in Figure 2. This spreads out the beam and makes it easier to see the interference. However, this spreading also means that only the central ray of the laser beam is still travelling on a straight line through the interferometer. All the surrounding rays are travelling at some angle, depending on how close to the center of the beam they are. Thus rays at different radii from the center of the laser beam travel a different total distance through the interferometer. This causes the interference pattern we see to look like a bullseye or target shape, with rings of bright and dark fringes instead of just one spot. During the experiment we will be counting bright-dark-bright fringe cycle. To do this you should pay attention to the center spot of the bullseye pattern, not to the outer part.

Since the two interfering beams of light were split from the same initial beam, they were initially in phase. Their relative phase when they meet at any point on the viewing screen, therefore, depends on the difference in the length of their optical paths in reaching that point. By moving mirror M 2 , the path length of one of the beams can be varied. Since the beam traverses the path between M 2 and the beam-splitter twice, moving M 2 one-quarter wavelength nearer the beam-splitter will reduce the optical path of that beam by one-half wavelength. The interference pattern will change; the radii of the maxima will be reduced so they

Figure 3: Interferometer

Suppose the micrometer knob is turned so it pushes the lever in by a distance d (see Figure 4). The angle of the lever arm changes by an amount θ such that d = R tan θ, as shown. Since the angle change is always small, R tan θ = Rθ, to a close approximation. This change in the lever arm angle causes the mylar strip to be pulled further around the lever post by an amount rθ, where r is the radius of the lever post. The mirror is therefore pulled away from the beam-splitter by the amount, rθ.

In this way, a relatively large displacement of the lever (d = Rθ) results in a much smaller displacement of the mirror (dm = rθ). By selecting appropriate values for r and R, the motion of M 2 is controlled so that each division on the micrometer dial corresponds to 1 micron of mirror movement.

Aligning the Interferometer

  1. Place the laser and the interferometer on the Optics Bench, approximately 10 - 20 cm apart (Figure 5). Be sure that the edges of both units are flush against the alignment rail of the bench. Place a viewing screen as shown. Turn on the laser.
  2. Loosen the thumbscrew that holds the beam-splitter and rotate the beam-splitter so it is out of the beam path of the laser as shown in Figure 5. Then loosen the thumbscrew that holds M 2 , the movable mirror. Adjust the rotation of M 2 so the laser beam is reflected directly back toward the aperture of the laser. (The reflected beam need not be at the same height as the incident beam, but it should strike the front panel of the laser along a vertical line through the aperture.) Hold M 2 in position and tighten the thumbscrew.
  3. Rotate the beam-splitter so its surface is at an angle approximately 45◦^ with the incident beam from the laser (see Figure 6). You will see two sets of laser spots on the viewing screen, corresponding to the two paths that the beam takes in reaching the screen. (Each path results in more than one laser spot because of multiple reflections within the beam-splitter.) Adjust the beam-splitter so

Figure 4: Mirror Movement Mechanism

the two sets of laser spots are as close as possible, then tighten the thumbscrew to secure the beam-splitter.

  1. Using the alignment screws, adjust the angle of M 1 until the two sets of laser spots are superimposed on the viewing screen (the two brightest spots must be superimposed).
  2. Place the lens holder on the optical bench as shown in Figure 7. Be sure its edge is flush against the alignment rail. Then place the 18 mm focal length lens on the lens holder (it attaches magnetically). Adjust the position of the lens on the holder so the light from the laser, now spread out by the lens, strikes the center of the beam-splitter. If you have performed the alignment correctly, you will see an interference pattern of concentric rings on the viewing screen. If the alignment is not just right, the center of the fringe pattern may not be visible on the screen. Adjust the alignment screws on M 1 very slowly as needed to center the pattern. This step requires patience, and may take you a while. The problem is that the beams along the two arms must be exactly parallel to properly interfere, and if one of the mirrors is tilted differently than the other, the beams wont be parallel. If you cant get the interference pattern, remove the lens and make sure the two brightest beams are still overlapped – by turning the mirror knobs too far you might have moved them apart. If they are not overlapped, fix them, replace the lens, and try again. You may need to move the lens up or down vertically by small amounts, or give the lens a slight twist to get the center part of the interference pattern displayed on the screen.
  3. Proceed once you have the interference pattern on the screen. Block the light along one arm. Observe what happens to the interference pattern, and explain.
  4. Tap one of the mirrors and observe what happens.

Figure 6: Adjusting M 1

movement. Remember, each division on the micrometer knob corresponds to one micron (10−^6 meters) of mirror movement. From these measurements calculate the wavelength of the laser light using

λ =

2 dm m

  1. Repeat the previous step measurements at least four more times for a total of five measurements of λ.
  2. Find the average measured value for λ and compare it to the accepted value for the HeNe laser wavelength, 632.8 nm. Do your average and the accepted value agree within the expected error?

PART B: Measuring the Index of Refraction for Air

For light of a specific frequency, the wavelength λ varies according to the formula:

λ =

λ 0 n

where λ 0 is the wavelength of the light in a vacuum, and n is the index of refraction for the material in which the light is propagating. In this experiment, you will use the interferometer to measure the index of refraction for air.

For reasonably low pressures, the index of refraction for a gas varies linearly with the gas pressure. Of course for a vacuum, where the pressure is zero, the index of refraction is exactly 1. A graph for the refraction index versus gas pressure is shown in Figure 8. The measurements you make in this experiment will allow you to calculate the slope of this graph for air. From that, numerical values can be determined for the index of air refraction at various pressures.

Figure 7: Positioning the Lens

  1. Align the laser and interferometer as described earlier.
  2. The experimental set-up is shown in Figure 9. Measure the length of the vacuum chamber. Push the air hose of the vacuum pump over the air outlet hole of the vacuum chamber. Then plug the banana plug of the vacuum chamber into the hole in the interferometer base — between the fixed mirror and the beam-splitter.
  3. Adjust the alignment screws of the fixed mirror so the center of the interference pattern is clearly visible on the viewing screen. (The fringe pattern will be somewhat distorted by irregularities in the glass end-plates of the vacuum chamber.)
  4. Notice that the banana plug of the vacuum chamber is free to rotate in its hole. For accurate measurements, the end-plates of the vacuum chamber must be perpendicular to the laser beam. Rotate the chamber slowly and observe the effect on the interference fringes. Based on your observations, how can you be sure that the end-plates of the vacuum chamber are properly aligned?
  5. Be sure that the air in the vacuum chamber is at atmospheric pressure. This is performed by flipping the vacuum release toggle switch. (Alternatively, many people find it more convenient to begin with the vacuum chamber evacuated — they then let air into the chamber as the fringes are counted. Use whichever method you find to be easier for controlling the air flow.)
  6. Record Pi, the initial reading on the vacuum pump gauge.

Figure 9: Experiment Setup for Index of Refraction of Air Measurement

pressure graph is therefore calculated as:

ni − nf Pi − Pf

∆mλ 0 / 2 d Pi − Pf where Pi = the initial air pressure, Pf = the final air pressure, ni = the index of refraction of air at pressure Pi, nf = the index of refraction of air at pressure Pf , ∆m = the number of fringes that passed the reference point during evacuation, λ 0 = the wavelength of the laser light in a vacuu, d = the length of the vacuum chamber (3.0 cm). Calculate the slope of the n vs pressure graph using the above.

  1. Repeat the previous Steps 6 – 8 for a minimum of ten trials and get an average value for the slope of the n vs pressure graph.
  2. Using your average slope, plot n vs P (you know that for P = 0, the index of refraction for the vacuum is n ≡ 1, so n = (slope) × P + 1).
  3. Using n = (slope) × P + 1, what is natm, the index of refraction of air at a pressure of 1 atmosphere (76.0 mm Hg). For the wavelength of HeNe light and at room temperature, the accepted value for the index of refraction of air is 1.00028. Is your experimentally determined value for natm within the estimated error of the accepted value?